Validation HW » MITS Systems » MITS-HFR
MITS-HFR

MITS-HFR

MITS High E-Field Resonators for Tier 1 and Tier 2 Evaluation at 64 MHz and 128 MHz

To support simplified Tier 1 and Tier 2 testing of implantable medical devices, ZMT’s Medical Implant Test System (MITS) can be complemented with targeted high electric (E-) field strength resonators (HFR). The MITS-HFR system enables evaluation of short implants (<100mm) directly at the maximum peak and rms E- field strengths without scaling and/or implant model generation and validation.

MITS High Field Resonator for 1.5T and 3.0T mounted on the MPE-AIMD Platform for measurements with the DASY6 AIMD scanner.

MITS-HFR1.5 and MITS-HFR3.0 generate the extremely high E- field exposures required for Tier 1 and Tier 2 evaluations as defined in the ISO/IEC standard TS 10974 Ed.2. The test conditions needed to allow access to these simplified assessments cannot typically be met using any other test environment. ZMT is unique in providing equipment for these field strengths for both 1.5T and 3T.

The resonators can be integrated with any existing MITS System. MITS-HFR1.5 and MITS-HFR3.0 have been rigorously validated. Each system can be used as a stand-alone or with a DASY6 AIMD scanner for rapid and high precision specific absorption rate (SAR) and temperature evaluation.

 

MITS-HFR1.5

 

Application

EMC, power deposition and heating testing of medical implants in maximum exposure conditions (e.g., according to Tier 1 & 2 of ISO/IEC standard TS 10974 Ed.2). Rapid, high-precision SAR and Temperature evaluations with DASY52NEO / DASY6 AIMD.
Evaluation of new prototypes

Operating frequency

64 MHz

Maximum induced E1

> 4,500 V/m (> 6,000 V/m with latest system with 18 KW peak power amplifiers) Tier 1 or Tier 2 Head and Trunk

Length of resonator

700 mm

Phantom Dimensions

300 x 100 x 100 mm

Weight of resonators

14.3 kg (17.3 kg with TSM)

DASY6 AIMD

Optimized for usage with DASY6 AIMD mounted in the MPE-AIMD platform

 

MITS-HFR3.0

 

Application

EMC, power deposition and heating testing of medical implants in maximum exposure conditions (e.g., according to Tier 1 & 2 of ISO/IEC standard TS 10974 Ed.2). Rapid, high-precision SAR and Temperature evaluations with DASY52NEO / DASY6 AIMD.
Evaluation of new prototypes

Operating frequency

128 MHz

Maximum induced E1

> 7,000 V/m

Length of resonator

450 mm

Phantom Dimensions

300 x 100 x 100 mm

Weight of resonators

11.7 kg (14.7 kg with TSM)

DASY6 AIMD